화학공학소재연구정보센터
Propellants Explosives Pyrotechnics, Vol.37, No.1, 116-121, 2012
Calibration and Uncertainty Analysis of an Optical Emission Spectrometer Measuring the Absolute Spectral Radiant Exitance of UV Signatures
An optical emission spectrometer (OES) was designed and constructed to detect the absolute spectral radiant exitance of UV signatures for a wavelength range of 240 nm to 440 nm. Using a directly heated graphite black body furnace operating at a temperature of 3000 K, the spectral response of the OES in the UV range was calibrated. To evaluate the performance of the OES system, the emission spectra of molecules were measured in a butane flame and the absolute emissions of the UV signatures were calculated. The uncertainty of the absolute measurement of the UV signature was analyzed by considering various uncertainty budgets, and the total uncertainty of the OES was found to be 1.66?%.