화학공학소재연구정보센터
Electrophoresis, Vol.34, No.11, 1563-1571, 2013
Determination of carbohydrate-deficient transferrin in human serum by capillary zone electrophoresis: Evaluation of assay performance and quality assurance over a 10-year period in the routine arena
The performance of high-resolution CZE for determination of carbohydrate-deficient transferrin (CDT) in human serum based on internal and external quality data gathered over a 10-year period is reported. The assay comprises mixing of serum with a Fe(III) ion-containing solution prior to analysis of the iron saturated mixture in a dynamically double-coated capillary using a commercial buffer at alkaline pH. CDT values obtained with a human serum of a healthy individual and commercial quality control sera are shown to vary less than 10%. Values of a control from a specific lot were found to slowly decrease as function of time (less than 10% per year). Furthermore, due to unknown reasons, gradual changes in the monitored pattern around pentasialo-transferrin were detected, which limit the use of commercial control sera of the same lot to less than 2 years. Analysis of external quality control sera revealed correct classification of the samples over the entire 10-year period. Data obtained compare well with those of HPLC and CZE assays of other laboratories. The data gathered over a 10-year period demonstrate the robustness of the high-resolution CZE assay. This is the first account of a CZE-based CDT assay with complete internal and external quality assessment over an extended time period.