화학공학소재연구정보센터
Journal of Colloid and Interface Science, Vol.416, 101-104, 2014
An induced current method for measuring zeta potential of electrolyte solution-air interface
This paper reports a novel and very simple method for measuring the zeta potential of electrolyte solution air interface. When a measuring electrode contacts the electrolyte solution air interface, an electrical current will be generated due to the potential difference between the electrode air surface and the electrolyte solution air interface. The amplitude of the measured electric signal is linearly proportional to this potential difference; and depends only on the zeta potential at the electrolyte solution air interface, regardless of the types and concentrations of the electrolyte. A correlation between the zeta potential and the measured voltage signal is obtained based on the experimental data. Using this equation, the zeta potential of any electrolyte solution air interface can be evaluated quickly and easily by inserting an electrode through the electrolyte solution air interface and measuring the electrical signal amplitude. This method was verified by comparing the obtained results of NaCl, MgCl2 and CaCl2 solutions of different pH values and concentrations with the zeta potential data reported in the published journal papers. (C) 2013 Elsevier Inc. All rights reserved.