화학공학소재연구정보센터
Electrophoresis, Vol.41, No.10-11, 852-859, 2020
Precise measurement and control of the pressure-driven flows for microfluidic systems
The pressure-driven device is designed and the flow rates of the microfluidic systems can be supplied by the pressure-driven flows, which can significantly reduce the flow-rate fluctuations coming from the pump source. For pressure-driven flows, the flow rates of the fluids can be predicted by measuring the pressure drop along a polytetrafluoroethylene (PTFE) tubing. Especially, by varying the geometrical parameters of the PTFE tubing, the predicted flow rates of the fluids are compared with the experimental measurements, and the testing precision of the pressure-driven flows can be obtained. Meanwhile, the dynamic characteristics of the open-loop and closed-loop control pressure-driven device are comparatively studied. Particularly, a proportional and integral (PI) controller is integrated with the closed-loop control pressure-driven device, and the effects of the parameters of the PI controller on the dynamic characteristics of the pressure-driven devices are mainly discussed. Most importantly, by improving the dynamic characteristics of the pressure-driven devices, precise measurement and control of the pressure-driven flows can be achieved for microfluidic systems.