화학공학소재연구정보센터
Journal of Vacuum Science & Technology B, Vol.11, No.6, 2137-2138, 1993
Temperature-Measurement for Scanning Tunnel Microscope Samples Using a Detachable Thermocouple
A detachable thermocouple contact for the measurement of sample temperature in ultrahigh vacuum was constructed and tested. The thermocouple was demonstrated to be accurate to within 1.2 K from 77 to 373 K. The thermocouple was also compared against a reference thermocouple of conventional design up to 1300 K during ohmic heating) and the agreement was within 20 K throughout this temperature range.