Thermochimica Acta, Vol.261, 183-194, 1995
Principles for the Interpretation of Modulated Temperature DSC Measurements .1. Glass-Transition
A new evaluation method for modulated temperature-DSC measurements based on linear response theory is presented. This yields a complex heat capacity with a real part (storage heat capacity) and an imaginary part (loss heat capacity). This approach makes irreversible (time dependent) thermal events amenable to quantitative and theoretically founded discussion. This is shown by theoretical analysis and demonstrated experimentally using the glass transition as an example.