Enzyme and Microbial Technology, Vol.26, No.7, 537-543, 2000
Screen-printed enzyme sensors for L-lysine determination
Sensors for the determination of L-lysine in samples of fermentation broth have been developed. Low-cost screen-printed sensors comprising a platinum working electrode, an Ag/AgCl pseudo reference and a carbon counter electrode were used as transducers for the enzyme sensors. L-lysine-(alpha)-oxidase from Trichoderma viride has been immobilized by entrapment into a polyurethane hydrogel. Sensors were characterized for L-lysine with respect to pH value, linear range, reproducibility, repeatability, storage and working stability. The sensitivities to other amino acids were also determined. A batch system with two working electrodes, one with immobilized enzyme and one without was adapted for the determination of L-lysine by differential measurements. Good agreement was found between L-lysine concentrations measured by the enzyme sensors and by a conventional amino acid analyzer.