화학공학소재연구정보센터
Biotechnology and Bioengineering, Vol.87, No.4, 473-477, 2004
Continuous glucose monitoring and control in a rotating wall perfused bioreactor
A glucose control system consisting of a single in-line glucose sensor, concentrated glucose solution, and computer hardware and software were developed. The system was applied to continuously control glucose concentrations of a perfusion medium in a rotating wall perfused vessel (RWPV) bioreactor culturing BHK-21 cells. The custom-made glucose sensor was based on a hydrogen peroxide electrode. The sensor continuously and accurately measured the glucose concentration of GTSF-2 medium in the RWPV bioreactor during cell culture. Three sets of two-point calibrations were applied to the glucose sensor during the 55-day cell culture. The system first controlled the glucose concentration in perfusing medium between 4.2 and 5.6 mM for 36 days and then at different glucose levels for 19 days. A stock solution with a high glucose concentration (266 mM) was used as the glucose injection solution. The standard error of prediction (SEP) for glucose measurement by the sensor, compared to measurement by the Beckman glucose analyzer, was +/-0.4 mM for 55 days. (C) 2004 Wiley Periodical, Inc.