Journal of Power Sources, Vol.146, No.1-2, 784-787, 2005
Performance and management of implantable lithium battery systems for left ventricular assist devices and total artificial hearts
A lithium ion cell designed for implantable medical devices was tested for its performance as a power source for left ventricular assist devices (LVAD) or total artificial hearts (TAH). These two cardiovascular devices require high power, and thus a high current (0.5-3 A) and high voltage (20-30 V). Since these are implantable medical devices, in addition to high power capability, the power source should have long cycle life and calendar life, as well as high safety. The QL0700I, a 700 mAh cell, was cycled at 0.5C rate as well as at 1.5C rate, and the cycle life capacity retention was evaluated after numerous cycles. A battery pack consisting of seven QL0700I cells in series, with a battery management system (BMS) connected, was tested for rate capability as well as safety protection. (c) 2005 Published by Elsevier B.V.