Electrochimica Acta, Vol.51, No.24, 5163-5168, 2006
A carboxyalkanethiol monolayer modified electrode for blocking of ascorbic acid oxidation and its application to a biosensor
Upon the application of amperometric biosensor to the biological fluid, ascorbic acid interferes the amperometric determination of analytes, because the oxidative potential of ascorbic acid is lower than that of electro active substances such as H2O2 produced by the enzymatic reaction. In this study we propose a method to block ascorbic acid based on the electrostatic interaction with self-assembled monolayer (SAM) and its application of the surface modified electrode to biosensor. In order to form SAM on the gold electrode with carboxyl group, 7-carboxy-heptanethiol (7-CHT) was used. The 7-CHT modified electrode did not show anodic response to ascorbic acid, but oxidized phenanthroline cobalt complex [Co(phen)(3)(2+)], which can be used as a mediator of biosensor. Thus, the 7CHT-modified electrode was applied to biosensor mediated with Co(phen)(3)(2+). Fructose dehydrogenase (FDH) was immobilized to the 7-CHT modified electrode. Fructose was determined selectively with the FDH/7-CHT modified electrode at the range of 0.2-2 mM. (c) 2006 Elsevier Ltd. All rights reserved.