Electrophoresis, Vol.21, No.8, 1620-1624, 2000
Preparation and characterization of calixarene-coated capillaries for capillary electrophoresis
Preparation and characterization of calixarene-coated capillaries for capillary electrophoresis (CE) were exemplified with p-allylcalix[4]arene (p-ACX(4)) which was immobilized to the fused silica surface using gamma-methacryloxypropyl-trimethoxysilane (gamma-MAPS) as linking agent. Successful p-ACX(4) coating was suggested by the greatly decreased electroosmotic flow (EOF), due to the introduction of phenolic hydroxyl groups on the inner surface of the capillaries. A slight slope of EOF versus pH at pH <8 would help make the separation reproducible. the coated columns also featured a low ultraviolet (UV) absorption background and long lifetime (> 6 months at 4