Fuel, Vol.77, No.14, 1527-1531, 1998
A test of the MALDI-MS calibration of SEC with N-methyl-2-pyrrolidinone for coal derived materials
In this paper we test our proposed methodology for a calibration procedure for mass determination by size exclusion chromatography. A sample is coarsely fractionated oil column and when the fractions are examined by matrix assisted laser desorption-ionisation mass spectrometry a direct variation of molecular mass with retention is found. Importantly the chromatograms produced when using N-methyl-2-pyrrolidinone as eluant are found to be free from solute association effects over a wide range of concentrations. The examination of a gasifier char is presented as a unique application of this chromatographic system.
Keywords:SIZE-EXCLUSION CHROMATOGRAPHY;GEL-PERMEATION CHROMATOGRAPHY;IONIZATION MASS-SPECTROMETRY;MOBILE-PHASE;LIQUEFACTION;PRODUCTS;LIQUIDS