Determination of Cd, Co, Ni and Pb in biological samples by microcolumn packed with black stone (Pierre noire) online coupled with ICP-OES
A simple and sensitive method using microcolumn (20-mm length x 2.0-mm i.d.) packed with black stone (Pierre noire) for the separation/preconcentration of Cd, Co, Ni and Pb in biological samples prior to their online determination by inductively coupled plasma optical emission spectrometry (ICP-OES) has been developed. Optimal experimental conditions including pH, eluent concentration and volume, sample volume and sample flow rate were investigated and established. The adsorption capacity of black stone for Cd, Co, Ni and Pb were found to be 23.4, 21.2, 18.1 and 22.2 mg g(-1), respectively. With a preconcentration time of 72 s and an elution time of 4 s, an enrichment factor of 20 and a sampling frequency of 25 h(-1) were obtained. The detection limits corresponding to three times the standard deviations of the blank for Cd, Co, Ni and Pb were found to be 0.3, 0.4, 0.4 and 1.1 ng mL(-1). The precision for seven replicate determinations of Cd, Co, Ni and Pb gave relative standard deviations (RSDs) of 5.9, 4.8, 2.7 and 1.1%, respectively (n=7, C= 10 ng mL(-1)). The method was validated with certified reference material GBW09103 human urine and the results obtained were in good agreement with the certified values. The method was also applied to the determination of the target analytes in biological samples with satisfactory results. (c) 2008 Elsevier B.V. All rights reserved.
Keywords:online microcolumn separation/preconcentration;heavy metals;black stone (Pierce noire);biological samples;ICP-OES