초록 |
The purpose of this study is to optimize the stent structure of D+storm DES through comparative evaluation with other stents. We evaluated 3 different commercially available Drug-eluting stent(DES) and D+storm DES. The radial force was higher in D product. The dislodgement force of the B, D products were higher than A, C products. There was no significant difference in the recoil and delivery force results. And the foreshortening results were observed in D product. The four cardiovascular stents has different physical properties. This will help you decide which stent is most suitable for clinical situation. This work was supported by the Korea Medical Device Development Fund grant funded by the Korea government (the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Ministry of Health & Welfare, the Ministry of Food and Drug safety)(Project Number: KMDF_PR_20200901_0246, 1711138528), (Project Number: KMDF_PR_20200901_0005, 9991006678) |