Thermochimica Acta, Vol.271, 127-140, 1996
Modulated Temperature DSC Measurements - The Influence of the Experimental Conditions
The influence of the experimental conditions on the results by modulated temperature DSC (MT-DSC) are investigated. The factors considered are temperature amplitude, frequency, underlying heating rates and sample preparation. Based on theoretical considerations and experimental results, practical recommendations are developed.